1
Using the 50-
Ω
feedthrough and the
BNC
cable, connect the signal
generator to channel 1 of the oscilloscope.
2
Set the frequency of the signal generator to 250 kHz and the
amplitude to about 800 mV.
3
Press
Autoscale
.
4
Adjust the output of the signal generator for exactly 8 divisions of
vertical deflection.
5
Change the frequency of the signal generator to the value shown
below for your instrument.
Table 4-6
Signal Generator Frequency Setting
Selected
Channel
HP 54600B
HP 54601B
HP 54602B *
HP 54603B
Channel 1
100 MHz
100 MHz
150 MHz
60 MHz
Channel 2
100 MHz
100 MHz
150 MHz
60 MHz
Channel 3
—
100
MHz
250
MHz
—
Channel 4
—
100
MHz
250
MHz
—
*1mv/div to 5 mv/div dc = 100 MHz.
6
Change the sweep speed of the oscilloscope to 5 ns/div and observe
the display.
The vertical amplitude of the signal on the display should be equal to or
greater than 5.66 divisions (
−
3 dB point). If the result is not
≤−
3 dB, see
"Troubleshooting the Oscilloscope," on page 4–40.
7
Repeat steps 1 through 6 for channel 2 (channels 2 to 4 on the
HP 54601B and HP 54602B).
Service
Verifying Oscilloscope Performance
4–11
Summary of Contents for HP 54600B
Page 7: ...vi ...
Page 12: ...1 The Oscilloscope at a Glance ...
Page 28: ...2 Operating Your Oscilloscope ...
Page 69: ...2 42 ...
Page 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Page 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Page 153: ...4 64 ...
Page 154: ...5 Performance Characteristics ...
Page 175: ...Glossary 8 ...