If the
Show Meas
softkey is turned on, cursors are displayed on the
waveform that show the measurement points for the right-most
measurement result. If you select more than one measurement, you
can show a previous measurement by reselecting the measurement.
•
To find the
Show Meas
softkey, press the
Next Menu
softkey key.
The oscilloscope makes automatic measurements on the first
displayed event. Figure 2-5 shows how to use delayed sweep to
isolate an event for a frequency measurement. If the measurement is
not possible in the delayed time base mode, then the main time base is
used. If the waveform is clipped, it may not be possible to make the
measurement.
Delayed time base isolates an event for a frequency measurement
Figure 2–5
Operating Your Oscilloscope
To make frequency measurements automatically
2–15