Keyboard
1
Remove the front panel.
2
Remove all the knobs by pulling straight out.
3
Flex the bezel of the front panel to unsnap the small keyboard under
the display opening.
4
Using the T10 TORX driver, remove the three screws from the large
keyboard.
Make sure that when you reinstall these screws that you use the correct
parts. If longer screws are used, they can damage the front-panel label.
5
Press down on the top of the keyboard, and rotate the bottom of the
keyboard out.
When installing the keyboard, make sure that the probe calibration cable is
kept away from the keyboard cable or noise can occur in the probe adjust
signal. See figure below for positioning the keyboard cable.
Positioning the keyboard cable
Probe calibration cable
Keyboard cable
Figure 4–10
Service
Replacing Parts in the Oscilloscope
4–55