To analyze video waveforms with Option 005
The combination of the TV trigger, delayed sweep, and automatic
measurements allow this oscilloscope to precisely analyze video waveforms.
There is no need for external clamping to obtain a stable trigger when you
are viewing unclamped video signals. This is because the TV sync separator
in the oscilloscope has an internal clamp circuit in the trigger path. Because
there is no clamp in the vertical path of your oscilloscope, you will be able to
observe any DC level shifts in the video on the oscilloscope display. To
eliminate this position shifting as the DC component of the video changes,
select AC coupling.
For this exercise, we connect the oscilloscope to the video output terminals
on a television. We set up the oscilloscope to view the second vertical
interval with delayed sweep windowed on the vertical interval test signals
(VITS). Then we make windowed measurements with the delayed sweep.
1
Connect a TV signal to channel 1, and select channel 1 as your trigger
source.
2
Press
Slope/Coupling
in the
TRIGGER
section of the front
panel, then press the
TV
softkey.
3
Select the desired TV
Standard
, such as
NTSC
,
PAL
,
PAL-M
or
SECAM
.
4
Press
Mode
, then press the
Video Autoscale
softkey.
Using Option 005 Enhanced TV/Video Trigger (HP 54602B)
To analyze video waveforms with Option 005
3–16
Summary of Contents for HP 54600B
Page 7: ...vi ...
Page 12: ...1 The Oscilloscope at a Glance ...
Page 28: ...2 Operating Your Oscilloscope ...
Page 69: ...2 42 ...
Page 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Page 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Page 153: ...4 64 ...
Page 154: ...5 Performance Characteristics ...
Page 175: ...Glossary 8 ...