Table 4-10
Equipment Required
Equipment
Critical specifications
Recommended
Model/Part
Signal generator
sine waves:
25 MHz and 100 MHz–HP 54600B/HP 54601B
25 MHz and 250 MHz–HP 54602B
25 MHz and 60 MHz–HP 54603B
Tek SG 503/Tek TM 501
Power splitter
(HP 54600B/03B only)
Outputs differ <0.15 dB
HP 11667B
Cable
BNC, Qty 3
HP 10503A
Adapter
N (m) to BNC (f), Qty 3
HP 1250-0780
Feedthrough
50
Ω
, BNC (m) and (f)
HP 10100C
Internal Trig
Sensitivity
1
Press
Setup
, then press the
Default Setup
softkey.
2
Connect the signal generator to channel 1.
3
Verify the trigger sensitivity at 25 MHz and 0.35 divisions.
a
Set the signal generator to 25 MHz and about 50 mV.
b
Press
Autoscale
.
c
Decrease the output of the signal generator until there is 0.35 vertical
divisions of the signal displayed.
The trigger should be stable. If the triggering is not stable, try
adjusting the trigger level. If adjusting the trigger level makes the
triggering stable, the test still passes. If adjusting the trigger does not
help, see "Troubleshooting the Oscilloscope," on page 4–40.
4
Verify the trigger sensitivity at 1 division for the frequencies shown
below.
a
Change the output of the signal generator to the frequency shown
below for your instrument and set amplitude to about 100 mV.
Service
Verifying Oscilloscope Performance
4–19
Summary of Contents for HP 54600B
Page 7: ...vi ...
Page 12: ...1 The Oscilloscope at a Glance ...
Page 28: ...2 Operating Your Oscilloscope ...
Page 69: ...2 42 ...
Page 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Page 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Page 153: ...4 64 ...
Page 154: ...5 Performance Characteristics ...
Page 175: ...Glossary 8 ...