To check out the oscilloscope
1
Is there an interface module connected to the oscilloscope?
If yes, do the following steps. If not, go to step 2.
a
Turn off the oscilloscope.
b
Remove the module.
c
Turn on the oscilloscope, then check for the failing symptom.
If the failing symptom disappears, replace the module. If not, go to
step 2.
2
Disconnect any external cables from the front panel.
3
Disconnect the power cord, then remove the cover.
4
Connect the power cord, then turn on the oscilloscope.
If the display comes on after a few seconds, (HP logo and copyright text,
followed by a graticule with text at top of the display) go to "To check the
Low Voltage Power Supply," on page 4–45. If after checking the Low Voltage
Power Supply the voltages are within the test limits, go to step 8. If not, go to
step 6. If the display did not come on, do the steps below.
a
Check the intensity knob setting to see if it
'
s set too low.
b
If there is still no display, disconnect the power cord.
c
Check all cable connections.
d
Go to "To check the Low Voltage Power Supply," on page 4–45.
If the voltages are within the limits go to step 5. If not, go to step 6.
Service
Troubleshooting the Oscilloscope
4–42
Summary of Contents for HP 54600B
Page 7: ...vi ...
Page 12: ...1 The Oscilloscope at a Glance ...
Page 28: ...2 Operating Your Oscilloscope ...
Page 69: ...2 42 ...
Page 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Page 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Page 153: ...4 64 ...
Page 154: ...5 Performance Characteristics ...
Page 175: ...Glossary 8 ...