Table 4-1
Recommended list of test equipment to service the oscilloscope
Equipment
Critical specifications
Recommended Model/Part
Use
Constant amplitude
signal
generator
100 MHz, Constant amplitude
±
1%
(250 MHz for HP 54602)
Tek SG503
Tek TM501
P
Digital multimeter
0.1 mV resolution, better than 0.01%
accuracy
HP 34401A
P, A, T
Oscilloscope
100 MHz
HP 54600B or equivalent
T
Power supply
14 mV to 35 Vdc, 0.1 mV resolution
HP 6114A
P
Probe
10:1 division ratio
HP 10432A
T
Pulse generator
Rise time < 875 ps
PSPL 1107B TD and PSPL 1110B Driver
A
Pulse generator
10 kHz, 500 mV p-p, rise time <5 ns
HP 8112A
A
Time marker
generator
Stability 5 ppm after 1/2 hour
TG 501A and TM 503A
P
Feedthrough
50
Ω
, BNC (m) and (f)
HP 10100C
P, A
Power splitter
Outputs differ <0.15 dB
HP 11667B
P
Shorting cap
BNC
HP 1250-0774
P
Adapter
SMA (f) to BNC (m)
HP 1250-1787
A
Adapter
BNC (f-f)
HP 1250-0080
A
Adapter
BNC tee (m) (f) (f)
HP 1250-0781
A
Adapter
N (m) to BNC (f), Qty 3
HP 1250-0780
P
Adapter
BNC (f) to dual banana (m)
HP 1251-2277
P
Cable
BNC, Qty 3
HP 10503A
P, A
Cable
BNC, 9 inches, Qty 2
HP 10502A
A
Additional equipment needed if you perform the alternate bandwidth test when verifying oscilloscope performance.
Signal generator
1 to 100 MHz at 200 mV
(250 MHz for HP 54602B)
HP 8656B opt 001
P
Power meter and
Power
Sensor
1 to 100 MHz
±
3% accuracy
(250 MHz for HP 54602B)
HP 436A and HP 8482A
P
Cable
Type N (m) 24 inch
HP 11500B
P
Adapter
Type N (m) to BNC (m)
HP 1251-0082
P
Service
4–3
Summary of Contents for HP 54600B
Page 7: ...vi ...
Page 12: ...1 The Oscilloscope at a Glance ...
Page 28: ...2 Operating Your Oscilloscope ...
Page 69: ...2 42 ...
Page 70: ...3 Using Option 005 Enhanced TV Video Trigger HP 54602B ...
Page 148: ...Exploded view of oscilloscope Figure 4 11 Service Replacing Parts in the Oscilloscope 4 59 ...
Page 153: ...4 64 ...
Page 154: ...5 Performance Characteristics ...
Page 175: ...Glossary 8 ...