98
Programming the Enhanced Monitor
Chapter 3
STATus:QUEStionable:BLOWer:LEVel?
The
STATus:QUEStionable:BLOWer:LEVel?
query returns an integer representing
the present fan level as a percentage of full scale speed. A percent symbol (%) is
attached to the value.
Returned Data
Comments
•
This query provides a percentage level of all the fans in the mainframe. It is not
a measurement. Individual fan speed measurements are provided by the
STATus:QUEStionable:BLOWer:SPEed?
query.
•
Example: on an RS-232 terminal, the command STAT:QUES:BLOW:LEV?
might return the string:
75%
Reset Condition
*RST does not affect the fan level directly. If the fans were on full speed because the
SYSTem:BLOWer:STATE FULL
command had been executed, then *RST clears
that state to VARiable.