162
Programming the Enhanced Monitor
Chapter 3
TEST:RESults[:CODE]?
TEST:RESults[:CODE]?
returns the results code for the last executed TEST, *TST,
or calibration command. The command returns an integer; refer to TEST:ALL? for
a complete list of test failure messages and CAL:ALL? for a complete list of
calibration failure commands.
Returned Data
Comments
•
The history queue also contains an entry for each test run and its results and
time stamp. Refer to
HIST:QUE:FETC?
•
Two other codes can be returned:
-- 478 indicating the last calibration or test was aborted.
-- 995 indicating that no calibration or test has been done since mainframe
power-on.
Reset Condition
*RST does not change the last test results.
Related
Commands
TEST:RES:VERB?
*TST?, TEST:ALL?, TEST:SENS?, TEST:TEMP?, TEST:DISP?, TEST:BLOW?,
TEST:TIME?, HIST:QUEue:FETCh?
Type
Range
Description
int16
-32768 to 32767
Result of test. +0 is a successful test. Refer to TEST:ALL? for a
complete list of test failure messages.