
Programming the Enhanced Monitor
105
Chapter 3
STATus:QUEStionable:CURRent:LIMit
<supply>,<value>
STATus:QUEStionable:CURRent:LIMit?
<supply> [,MIN|MAX]
The
STATus:QUEStionable:CURRent:LIMit
sets the limit for amperage on one of the
seven supplies in the mainframe. If it is exceeded, a warning is generated. These
values are stored in non-volatile memory.
The
STATus:QUEStionable:CURRent:LIMit?
returns a floating number representing
the limits for amperage in the mainframe.
Parameters
Returned Data
Comments
•
Power Supply Maximums:
•
This command corrects the polarity of <value> to match the specified
<supply>. It then determines if <value> is within the valid range for the
specified <supply>; if it is not, the limit is set to MAX. No error is generated.
•
Remember to execute a
SYSTem:NVSave
after setting all the limits you desire.
This will move them into non-volatile memory.
Reset Condition
*RST resets register the last value saved by the SYST:NVS command.
Related
Commands
SYSTem:NVSave
Name
Type
Range
Default
Description
<
supply>
enum.
P24, P12, P5, N2,
N5PT2, N12, N24
none
Indicates which power supply
current limit is being set. P5 is
positive 5 Vdc supply, N12 is
negative 12 Vdc supply, etc.
<
value>
float32 -1000 to 1000,
MIN, MAX
Min is either
+1.0 or -1.0;
see table below
for MAX
Current limit. Each supply has
its own maximum. Use
negative value for negative
supplies.
optional
parameter
enum.
MIN
MAX
Returns maximum or minimum
current limit.
Type
Range
Description
float
Current limit for <
supply>.
Power Supply
HP E8402
HP E8404
P5
50A
90A
P12
6A
15A
N12
– 4A
– 15A
P24
4A
15A
N24
– 4A
– 15A
N5PT2
– 20A
– 60A
N2
– 10A
– 30A
Summary of Contents for E8402A
Page 8: ...10 Contents ...
Page 12: ...14 Notes ...
Page 14: ......
Page 26: ...28 Getting Started Chapter 1 ...
Page 42: ...44 Using the Enhanced Monitor Chapter 2 ...
Page 214: ...216 Servicing Your Mainframe Chapter 5 ...
Page 226: ...228 HP E8402 E8404A Product Specifications Appendix A ...