70
Programming the Enhanced Monitor
Chapter 3
HISTory:POWer:MAXimum?
<
supply
>
HISTory:POWer:MAXimum?
<supply> returns a single floating point number for the
maximum wattage measured by the Enhanced Monitor for the <supply> since either
manufacture of the mainframe or the most recent
HIST:RES:POW
or
HIST:RES:ALL
.
Parameters
Returned Data
Comments
•
The maximum wattage tracking can be zeroed by the
HISTory:RESet
commands.
Error Conditions
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
SYSTem:ERRor?
query.
Reset Condition
A *RST has no effect on the HISTory subsystem.
Related
Commands
HISTory:POWer:CMAXimum?, HISTory: POWer:[HISTogram]?,
HISTory:RESet: POWer, HISTory:RESet[:ALL]
Name
Type
Range
Default
Description
<
supply>
enum
P5, P12, N12, P24,
N24, N5PT2, N2,
TOTal
none
Selects the power supply for the
maximum wattage request. P5 is
positive 5 Vdc supply, N12 is
negative 12 Vdc supply, etc.
Type
Range
Description
float
Maximum current (in amperes) for specified power supply.
Number
Message
Probable Cause
-109
“Missing parameter”
You must include the <
supply> parameter.
-221
“Settings Conflict”
The power supply doesn’t support power
measurements.
-224
“Illegal Parameter”
The
<supply> parameter was not correct.
-311
“Memory error”
The historical data’s memory is not available.