80
Programming the Enhanced Monitor
Chapter 3
HISTory:TIME:LHReset?
HISTory:TIME:LHReset?
returns the amount of operating time since the last
HISTory:RESet
(or any specific H
IST:RES:xxxx
) command.
Returned Data
Comments
•
If no
HISTory:RESet
has been performed, this query returns the amount of
time since mainframe manufacture. The history queue contains information on
which reset was done.
Error Conditions
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
SYSTem:ERRor?
query.
Reset Condition
*RST has no effect on this query.
Related
Commands
HISTory:RESet[:ALL], HISTory:RESet:TEMPerature, HISTory:RESet:QUEue
Type
Range
Default
Description
<
hours>
uint32
0 to 4294967295
none
hours since history reset
<
min>
uint16
0 to 60
none
+ minutes since history reset
<
sec>
uint16
0 to 60
none
+ seconds since history reset
Number
Message
Probable Cause
-311
“Memory Error”
Non-volatile memory is inaccessible.