Programming the Enhanced Monitor
67
Chapter 3
HISTory:CURRent:MAXimum?
<
supply
>
HISTory:CURRent:MAXimum?
<supply> returns a single floating point number for
the maximum amperage that has occurred to the <supply> since manufacture of the
mainframe, or the most recent
HIST:RES:CURR
or
HIST:RES:ALL
.
Parameters
Returned Data
Comments
•
The maximum amperage tracking can be zeroed by the
HISTory:RESet
commands.
Error Conditions
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
SYSTem:ERRor?
query.
Reset Condition
A *RST has no effect on the HISTory subsystem.
Related
Commands
HISTory:CURRent:CMAXimum?, HISTory:CURRent:[HISTogram]?,
HISTory:RESet:CURRent, HISTory:RESet[:ALL]
Name
Type
Range
Default
Description
<
supply>
enum
P5, P12, N12, P24,
N24, N5PT2, N2
none
Selects the power supply for the
maximum amperage request. P5 is
positive 5 Vdc supply, N12 is
negative 12 Vdc supply, etc.
Type
Range
Description
float
Maximum current (in amperes) for specified power supply.
Number
Message
Probable Cause
-221
“Settings Conflict”
The power supply doesn’t support current
measurements.
-224
“Illegal Parameter”
The
<supply> parameter was not correct.
-311
“Memory error”
The historical data’s memory is not available.
Summary of Contents for E8402A
Page 8: ...10 Contents ...
Page 12: ...14 Notes ...
Page 14: ......
Page 26: ...28 Getting Started Chapter 1 ...
Page 42: ...44 Using the Enhanced Monitor Chapter 2 ...
Page 214: ...216 Servicing Your Mainframe Chapter 5 ...
Page 226: ...228 HP E8402 E8404A Product Specifications Appendix A ...