116
Programming the Enhanced Monitor
Chapter 3
STATus:QUEStionable:UMCounter:TINTerval <
time
>
STATus:QUEStionable:UMCounter:TINTerval?
The
STATus:QUEStionable:UMCounter:TINTerval
command sets the time interval
value of the user maintenance counter. The user maintenance counter is a countdown
timer that counts hours down to zero. The time interval is the value from which the
UMCounter will start. This value is stored in non-volatile memory immediately.
The S
TATus:QUEStionable:UMCounter:TINTerval?
command returns an integer
representing the time interval value of the user maintenance counter.
Parameters
Returned Data
Comment
•
The value 65535 means that the counter is disabled. It is shipped from the
factory disabled. Set the time interval (
STAT:QUES:UMC:TINT
) and then reset
the counter (
STAT:QUES:UMC:TRES
) to enable the user maintenance counter.
To disable the counter, send the command
STATus:QUEStionable:UMCounter:TINTerval DEF
.
•
The MINimum value is 1, MAXimum value is 65534, DEFault value is 65535.
Reset Condition
*RST does not affect the UM Counter.
Related
Commands
STATus:QUEStionable:UMCounter:TRESet,
STATus:QUEStionable:UMCounter:TREMaining?
Name
Type
Range
Default
Description
<
time>
int16
1 to
65535,MIN,
MAX, DEF
65535
number of hours the UMCounter will count
down from.
Type
Range
Description
int16
User Maintenance Counter