166
Programming the Enhanced Monitor
Chapter 3
TEST:TIME?
TEST:TIME?
performs a test of the timing functionality of the enhanced monitor.
This test outputs a waveform on Pin 25 of the Diagnostic Connector with a 16 msec
period for 10 seconds. The waveform will start within 2 seconds of the command.
This waveform can be measured with a counter to determine its accuracy and jitter.
There is no failure return for this test.
Returned Data
This command always returns a +0. Refer to Chapter 4 for instructions on
performing this test.
Comments
•
TEST:TIME?
interrupts the monitoring function of the enhanced monitor.
Measurement levels returned from other commands during a test routine were
all measured before the test routine started.
Error Conditions
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using
SYSTem:ERRor?
query.
Reset Condition
*RST aborts any TEST being performed.
Related
Commands
*TST?, TEST:ALL?, TEST:SENS?, TEST:TEMP?, TEST:DISP?, TEST:BLOW?,
TEST:MEM?
Number
Message
Probable Cause
-221
“Settings Conflict”
The mainframe’s power is off.
Summary of Contents for E8402A