110
Programming the Enhanced Monitor
Chapter 3
STATus:QUEStionable:TEMPerature:CONDition?
The
STATus:QUEStionable:TEMPerature:CONDition?
query returns an integer
representing the contents of the condition register associated with the TEMPerature
status group.
The condition register continuously monitors the hardware and firmware status of
the instrument. There is no latching or buffering for this register; it is updated in real
time. Condition registers are read-only.
The following table describes each bit in the TEMPerature status group:
Returned Data
Comments
A bit is set when a slot temperature is over the user-specified limit.
Reset Condition
*RST does not affect the status system.
Related
Commands
STATus:QUEStionable:TEMPerature:EVENt?,
STATus:QUEStionable:TEMPerature:ENABle,
STATus:QUEStionable:TEMPerature:LIMit
Bit
Description
Decimal Value
0
Slot 0
1
1
Slot 1
2
2
Slot 2
4
3
Slot 3
8
4
Slot 4
16
5
Slot 5
32
6
Slot 6
64
7
Slot 7
128
8
Slot 8
256
9
Slot 9
512
10
Slot 10
1024
11
Slot 11
2048
12
Slot 12
4096
13
AMBient
8192
14
Power supply
16384
15
Unused
32768
Type
Range
Default
Description
int16
0 - 32767 none
Bit mask indicating which conditions are set (1) or cleared (0)
for the TEMPerature status group.
Summary of Contents for E8402A
Page 8: ...10 Contents ...
Page 12: ...14 Notes ...
Page 14: ......
Page 26: ...28 Getting Started Chapter 1 ...
Page 42: ...44 Using the Enhanced Monitor Chapter 2 ...
Page 214: ...216 Servicing Your Mainframe Chapter 5 ...
Page 226: ...228 HP E8402 E8404A Product Specifications Appendix A ...