182
Programming the Enhanced Monitor
Chapter 3
*TST?
*TST? performs several tests of the enhanced monitor functionality. This is a subset
of the full test set TEST:ALL? The tests performed are TEST:MEMory,
TEST:SENSe, TEST:TEMPerature, and TEST:DISPLAY.
Returned Data
Comments
•
*TST? interrupts the monitoring function of the enhanced monitor for four
seconds during the end of the test.
•
Refer to
TEST:ALL
for a complete list of test failure results.
Error Conditions
The following table lists the most common error conditions and causes. Error
numbers and corresponding messages can be found using SYSTem:ERRor? query.
Related
Commands
CALibration?
Type
Range
Default
Description
int16
-32767 to 32767
none
Result of running the self test.
+0 indicates a successful test. Refer to TEST:ALL for a
complete list of test failure results.
Number
Message
Probable Cause
-221
“Settings Conflict”
the mainframe’s power is off.