Programming the Enhanced Monitor
123
Chapter 3
STATus:QUEStionable:VOLTage:PTR <
mask
>
STATus:QUEStionable:VOLTage:PTR?
The
STATus:QUEStionable:VOLTage:PTR
sets the voltage positive transition filter.
The
STATus:QUEStionable:VOLTage:PTR?
returns an integer representing the
voltage positive transition filter.
Only bit 3 and bit 4 of the transition filter can be set or cleared by the user.
Parameters
Returned Data
Comments
•
If *PSC was set to 0 prior to power-down, then at power-on, the enable
registers are loaded from non-volatile memory. Therefore, any values that you
have stored (
SYSTem:NVSave
command) are returned to the registers.
However, if *PSC was set to 1, then at power-on the factory defaults are
returned.
•
The following table describes the bits in the VOLTage status group:
Name
Type
Range
Default
Description
<
mask> int16
0 to 32767
511
(see comment
below)
Indicates the transition direction when an
event will occur.
Type
Range
Default
Description
int16
0 to 32767
none
Indicates the transition direction when an event will occur.
Bit
Description
Decimal Value
0
+24 (P24) volt power supply
1
1
+12 (P12) volt power supply
2
2
+5 (P5) volt power supply
4
3
5Volt Standby
8
4
External 5 volt supply for
Enhanced Monitor RS-232 port.
16
5
-2 (N2) volt power supply
32
6
-5.2 (N5P2) volt power supply
64
7
-12 (N12) volt power supply
128
8
-24 (N24) volt power supply
256
9
Unused 512
10
Unused
1024
11
Unused
2048
12
Unused
4096
13
Unused
8192
14
Unused
16384
15
Unused
32768