154
Programming the Enhanced Monitor
Chapter 3
SYSTem:POWer:STATus?
SYSTem:POWer:STATus?
returns three integers representing the status of the three
sources of power-up/power-down assertion. The three sources are the front panel
ON/STDBY switch, the diagnostic connector’s remote power lines, and the
SYSTem:POWer command.
Returned Data
Comments
The mainframe cannot power-off unless ALL three of these states are 0 (zero).
Reset Condition
*RST has no effect on this query.
Related
Commands
SYSTem:POWer
Type
Range
Default
Description
<switch>
unit16n
0, 1
none
This indicates the front panel switch state. 1
indicates it is pushed in (power-on).
<
auxRemote>
unit16
0, 1
none
The state of the Diagnostic connector’s
remote power signal. 1 indicates it is
connected to the remote power signal return.
Refer to “Using the Remote Power-On Pins”
on page 25 and “Diagnostic Connector” on
page 42
<
software>
unit16
0, 1
none
The state of the software power-on
command (SYSTem:POW ON/OFF). 1
indicates that a software power-on command
is active.
Summary of Contents for E8402A