order:
0440
0440,"Query
UNTERMINA
TED
after
indenite
resp onse;(
0440)"
Correct
the
HP-IB
con
troller
program
so
that
the
query
that
returns
indenite
length
blo
c
k
data
is
the
last
item
on
the
program
line.
0430
0430,"Query
DEADLOCKED;(
0430)"
Correct
the
HP-IB
con
troller
program
so
that
no
more
than
eigh
t
queries
are
executed
within
the
same
line
of
the
program.
0420
0420,"Query
UNTERMINA
TED;( 0420)"
Correct
the
HP-IB
con
troller
program
so
that
the
con
troller
terminates
commands
with
the
newline
c
haracter
(NL)
b efore
the
con
troller
attempts
to
read
query
resp onse
data.
0410
0410,"Query
INTERRUPTED;(0410)"
Chec
k
the
HP-IB
con
troller
program
to
see
if
the
con
troller
is
programmed
to
read
the
en
tire
query
resp onse
data
b efore
issuing
a
subsequen
t
command.
0400
0400,"Query
error;(0400)"
Some
problem
o ccurred
while
parsing
an
HP-IB
query
.
Insure
that
y
our
programming
is
correct
and
try
the
query
again.
Lo ok
at
0440
through
0400
for
t
yp es
of
problems
to
lo
ok
for.
0350
0350,"Queue
o
v
ero
w"
The
error
queue
o
v
ero
w
ed
at
this
p oint
and
this
message
replaced
the
16th
error
message.
No
action
is
required.
Note:
T
o
clear
the
HP-IB
error
queue,
use
*CLS.
0310
0310,"System
error;(0310)"
Some
problem
o ccurred
while
parsing
an
HP-IB
command
or
query
.
Insure
that
y
our
programming
is
correct
and
try
the
command
again.
0278
0278,"Macro
header
not
found;( 0278)"
A
*GMC?
or
*RMC
macro
lab
el
could
not
b e
found
in
the
list
of
dened
macro
lab
els.
Use
*LMC?
to
get
a
list
of
all
the
curren
tly
dened
macro
lab
els.
0277
0277,"Macro
redenition
not
allow
ed;(
0277)"
Indicates
that
a
macro
lab
el
in
the
*DMC
command
could
no
b e
dened
b ecause
the
macro
lab
el
w
as
already
dened.
0276
0276,"Macro
recursion
error;(0276)"
The
nesting/recursion
of
macros
got
deeper
than
4
lev
els.
Don't
use
more
than
4
lev
els
when
dening
macros
of
macros.
0275
0275,"Macro
denition
to
o
long;(
0275)"
The
macro
denition
m
ust
b e
255
c
haracters
or
less.
0274
0274,"Macro
parameter
error;(0274)"
A
macro
parameter
placeholder
w
as
improp
erly
used.
0273
0273,"Illegal
macro
lab el;( 0273)"
Error
Messages
8-3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for 70340A
Page 94: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 110: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 224: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 402: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 424: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 436: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...