11-5
MAINTENANCE PERIPHERALS
• Reinsert the piston into the syringe.
• Reinsert the syringe by screwing it in the valve with your fingers.
• Reinsert the protective cap to the bottom of the burette assembly. Carefully position the cap into the burette.
• Slide the burette into the burette stand. Notice the position of the piston shaft to the pump couple.
• Priming the burette three times with new titrant is recommended.
11.1.6. BURETTE PREPARATION (TITRANT FILLING)
Before starting a titration, the burette must be properly filled with titrant in order to obtain an accurate and repeatable result. To fill
the burette, follow the next steps and recommendations:
• If necessary, clean the burette and make sure it is empty.
• From the main screen press
.
• Highlight
Prime Burette
option and press
.
• Enter the number of times the burette needs to be rinsed (minimum three rinses are recommend allowing air bubbles to be
evacuated).
• Press
.
To avoid the presence of the air bubbles inside the burette, make sure to have a continuous liquid flow inside the burette. A little air
just above the liquid level at the first filling is normal. The next filling will evacuate all of the air; no air will be left in the valve.
Sometimes during this process, slight finger tapping on the tubes is helpful to remove any residual air bubbles from the tubes.
If air bubbles are still present:
• Remove the aspiration tube from the titrant bottle.
• Repeat burette preparation procedure.
• If this is not successful, clean the burette again.
Summary of Contents for HI931
Page 1: ...MANUALS AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 2: ......
Page 4: ......
Page 5: ...QUICK START GUIDE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 9: ...5 QUICK START GUIDE TITRATOR CONNECTIONS FRONT VIEW REAR VIEW ...
Page 19: ...INSTRUCTION MANUAL AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 20: ......
Page 24: ...1 2 ...
Page 26: ...2 2 ...
Page 29: ...2 5 SETUP 2 3 2 TITRATOR REAR VIEW 2 3 3 TITRATOR REAR VIEW WITH PERISTALTIC PUMP ...
Page 36: ...2 12 ...
Page 38: ...3 2 ...
Page 46: ...4 2 ...
Page 92: ...5 34 ...
Page 94: ...6 2 ...
Page 98: ...6 6 ...
Page 100: ...7 2 ...
Page 112: ...7 14 ...
Page 114: ...8 2 ...
Page 120: ...8 8 ...
Page 122: ...9 2 ...
Page 134: ...9 14 ...
Page 136: ...10 2 ...
Page 144: ...10 10 ...
Page 146: ...11 2 ...
Page 154: ...A1 2 ...
Page 158: ...A2 2 ...
Page 169: ...GENERAL TITRATION APPLICATIONS BROCHURE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 170: ......
Page 195: ...TITRATION THEORY AUTOMATIC POTENTIOMETRIC TITRATOR HI932 HI931 ...
Page 196: ...2 ...
Page 198: ...4 ...
Page 216: ...22 THEORY931 932 10 18 ...
Page 217: ......