7-11
pH MODE
PREPARATION
Pour small quantities of the buffer solutions into clean beakers. If possible, use plastic beakers to minimize any EMC interferences.
For accurate calibration and to minimize cross-contamination, use two beakers for each buffer solution: one for rinsing the electrode
and one for calibration. If you are measuring in the acidic range, use pH 7.01 or 6.86 as the first buffer and pH 4.01/3.00 or 1.68
as the second buffer. If you are measuring in the alkaline range, use pH 7.01 or 6.86 as the first buffer and pH 10.01/9.18 or
12.45 as the second buffer.
For extended range measurements (acidic and alkaline), perform a five-point calibration by selecting five buffers across the entire pH
range.
CALIBRATION PROCEDURE
During calibration, the user has a choice of 8 standard buffers: (pH 1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) and up to 5
custom buffers.
For accurate measurements it is recommended to perform a five-point calibration. However, at least a two-point calibration is
suggested. For pH titrations, the selected buffers should bracket your end point (e.g.: if your end point value is at 8.5, use 7.01 or
6.86 and 9.18 or 10.01 for calibration).
To begin calibration:
• Press
. If the instrument was calibrated before the old calibration can be cleared by pressing
.
Note:
It is very important to clear calibration history when a new electrode is used.
• Immerse the pH electrode and the temperature probe approximately 4 cm (1.5”) into a buffer solution and stir gently.
• If necessary, select the pH calibration buffer value with
or
.
• Once the reading has stabilized press
to update the calibration. The calibration buffer will be added to the Calibrated
Buffers section.
• Rinse the pH electrode and the temperature probe, then immerse them into the next buffer solution and follow the above
procedure or press
to exit the calibration.
Notes:
•
The new calibration points will replace old ones if the difference between them is ± 0.2 pH.
•
Buffers used in older calibrations will not have a solid background.
•
If calibrating with a standard buffer in MTC mode, the pH value and temperature can be modified by pressing
. The
values can be adjusted using the numeric keys. Press
to save the new values.
Summary of Contents for HI931
Page 1: ...MANUALS AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 2: ......
Page 4: ......
Page 5: ...QUICK START GUIDE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 9: ...5 QUICK START GUIDE TITRATOR CONNECTIONS FRONT VIEW REAR VIEW ...
Page 19: ...INSTRUCTION MANUAL AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 20: ......
Page 24: ...1 2 ...
Page 26: ...2 2 ...
Page 29: ...2 5 SETUP 2 3 2 TITRATOR REAR VIEW 2 3 3 TITRATOR REAR VIEW WITH PERISTALTIC PUMP ...
Page 36: ...2 12 ...
Page 38: ...3 2 ...
Page 46: ...4 2 ...
Page 92: ...5 34 ...
Page 94: ...6 2 ...
Page 98: ...6 6 ...
Page 100: ...7 2 ...
Page 112: ...7 14 ...
Page 114: ...8 2 ...
Page 120: ...8 8 ...
Page 122: ...9 2 ...
Page 134: ...9 14 ...
Page 136: ...10 2 ...
Page 144: ...10 10 ...
Page 146: ...11 2 ...
Page 154: ...A1 2 ...
Page 158: ...A2 2 ...
Page 169: ...GENERAL TITRATION APPLICATIONS BROCHURE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 170: ......
Page 195: ...TITRATION THEORY AUTOMATIC POTENTIOMETRIC TITRATOR HI932 HI931 ...
Page 196: ...2 ...
Page 198: ...4 ...
Page 216: ...22 THEORY931 932 10 18 ...
Page 217: ......