22
ALTERNATIVE CALCULATIONS
If the actual values of the above parameters are not accessible the
following equation can be used:
V=M*F
V Volume of measured mass of water (mL)
F Transformation factor
The transformation factor takes into account the air buoyancy, the
water density and temperature dependence. Standard values can be
used to obtain the transformation factor.
The values from the table below have been calculated by correcting
the air and water density with temperature, assuming the density of
dry air
ρ
air
= 0.0012 g/mL and density of calibration steel standard
weigh
ρ
STD
= 8 g/mL.
Temperature ( °C)
Factor
17.0
1.002290
18.0
1.002467
19.0
1.002654
20.0
1.002853
21.0
1.003061
22.0
1.003282
23.0
1.003512
24.0
1.003752
25.0
1.004002
26.0
1.004261
27.0
1.004531
28.0
1.004809
29.0
1.005097
30.0
1.005395
TEMPERATURE CHANNEL FAST CHECK PROCEDURE
• Connect the shorting cap to the BNC socket on Analog Board 1.
• Connect the HI762000C 0°C temperature key to the RCA socket
(temperature sensor input) on Analog Board 1.
• On the main screen select
, if necessary select the
analog board and press
.
• The titrator should display ATC 0.0 ± 0.4°C with no fluctuations
or drift.
• Connect the HI762070C 70°C temperature key to the RCA socket
(temperature sensor input) on Analog Board 1.
• The titrator should display ATC 70.0 ± 0.4°C with no
fluctuations or drift.
• This procedure can be repeated on analog board 2.
TEMPERATURE & MV CHANNEL LOGGING PROCEDURE
• Connect the shorting cap to the BNC socket on Analog Board 1.
• Connect the HI762000C 0°C temperature key to the RCA socket
(temperature sensor input) on Analog Board 1.
• On the main screen select
, if necessary select the
analog board and press
.
• Press
and use the arrow keys to highlight
Logging
Interval
. Set the logging interval to 15 seconds and press
. Press
to return to the main screen.
• Press the
key and use the arrow keys to highlight
Setup
pH/mV/ISE Report
, press
.
• Select
Potential and Temperature and Units
(the selected fields
are marked with an *). All other fields should be unselected.
• Press
to return to the Data Parameters screen.
• Press
to return to the main screen.
• Once on the main screen press
to start the automatic
log.
• Let the log run for about 10 minutes. Press
to stop
the automatic log.
• Press
, use the arrow keys to highlight
Review Last
Analysis Report
, and press
.
• The mV column should display 0.0 ± 0.1 mV and the
temperature column should display 0.0°C ± 0.4°C.
• This procedure can be repeated using the HI762070C 70°C
temperature key and on analog board 2.
TROUBLESHOO
TING 1
TROUBLESHOOTING 1
METHOD ID:
HI1011EN
Summary of Contents for HI931
Page 1: ...MANUALS AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 2: ......
Page 4: ......
Page 5: ...QUICK START GUIDE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 9: ...5 QUICK START GUIDE TITRATOR CONNECTIONS FRONT VIEW REAR VIEW ...
Page 19: ...INSTRUCTION MANUAL AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 20: ......
Page 24: ...1 2 ...
Page 26: ...2 2 ...
Page 29: ...2 5 SETUP 2 3 2 TITRATOR REAR VIEW 2 3 3 TITRATOR REAR VIEW WITH PERISTALTIC PUMP ...
Page 36: ...2 12 ...
Page 38: ...3 2 ...
Page 46: ...4 2 ...
Page 92: ...5 34 ...
Page 94: ...6 2 ...
Page 98: ...6 6 ...
Page 100: ...7 2 ...
Page 112: ...7 14 ...
Page 114: ...8 2 ...
Page 120: ...8 8 ...
Page 122: ...9 2 ...
Page 134: ...9 14 ...
Page 136: ...10 2 ...
Page 144: ...10 10 ...
Page 146: ...11 2 ...
Page 154: ...A1 2 ...
Page 158: ...A2 2 ...
Page 169: ...GENERAL TITRATION APPLICATIONS BROCHURE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 170: ......
Page 195: ...TITRATION THEORY AUTOMATIC POTENTIOMETRIC TITRATOR HI932 HI931 ...
Page 196: ...2 ...
Page 198: ...4 ...
Page 216: ...22 THEORY931 932 10 18 ...
Page 217: ......