13
ACIDITY OF W
ATER
ACIDITY OF WATER
0 to 2500 mg/L, pH 8.3 Endpoint
METHOD ID:
HI1005EN
DESCRIPTION
Method for the determination of total (phenolphthalein) acidity in
water by titration of a sample to pH 8.3. The results are expressed in
mg/L (ppm) as calcium carbonate
.
For the determination of methyl orange acidity, set the endpoint to
pH 3.7.
REFERENCE
Standard Methods for the Examination of Water and Wastewater
21
st
edition, Method 2310B
ELECTRODE
• HI1131B Combination pH Electrode
• HI7662-T Temperature Probe
REAGENTS
• HI70456 0.1N Sodium Hydroxide (1 L)
• HI70436 Deionized Water (1 gal)
ACCESSORIES
• HI70300L Storage Solution (500 mL)
• HI7082
Electrode Fill Solution (4 x 30 mL)
• HI7004L pH 4.01 Buffer Solution (500 mL)
• HI7007L pH 7.01 Buffer Solution (500 mL)
• HI7010L pH 10.01 Buffer Solution (500 mL)
• HI740036P 100 mL Plastic Beaker (10 pcs)
• 50 mL Class A Volumetric Pipette
DEVICE PREPARATION
• Connect the pH electrode and temperature probe to the titrator.
• Install a 25 mL burette filled with 0.1N sodium hydroxide
(HI70456) on pump one and verify that no air bubbles are
present in the burette or tubing. If necessary prime the burette
until all the air has been removed completely.
• For the determination of the exact concentration of the 0.1N
sodium hydroxide, follow
HI0001EN 0.1N Sodium Hydroxide
Titrant Concentration.
• Press
from the main screen. Use the arrow keys to
highlight
Acidity of Water
and press
.
ELECTRODE PREPARATION
• Press
from the main screen, if necessary select the
analog board and press
.
• Calibrate the electrode using pH 4.01, 7.01 and 10.01 buffers.
Refer to the instruction manual for calibration procedure.
SAMPLE PREPARATION
• Use a Class A volumetric pipette to transfer exactly 50.00 mL of
sample to a clean 100 mL plastic beaker.
ANALYSIS
• Place the beaker under the stirrer assembly and lower it to
immerse the pH electrode, temperature sensor and stirrer.
Ensure that the reference junction of the pH electrode is 5 to 6
mm below the surface. If necessary add extra deionized water.
Note
:
The dispensing tip should be slightly submerged in the
sample.
• Press
, the titrator will start the analysis.
• At the end of the titration, when pH 8.30 is reached, “Titration
Completed” will appear with the result. The result is expressed
in
mg/L as calcium carbonate.
• Remove the pH electrode, temperature probe and stirrer from the
sample and rinse them thoroughly with deionized water.
• Record the result.
Summary of Contents for HI931
Page 1: ...MANUALS AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 2: ......
Page 4: ......
Page 5: ...QUICK START GUIDE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 9: ...5 QUICK START GUIDE TITRATOR CONNECTIONS FRONT VIEW REAR VIEW ...
Page 19: ...INSTRUCTION MANUAL AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 20: ......
Page 24: ...1 2 ...
Page 26: ...2 2 ...
Page 29: ...2 5 SETUP 2 3 2 TITRATOR REAR VIEW 2 3 3 TITRATOR REAR VIEW WITH PERISTALTIC PUMP ...
Page 36: ...2 12 ...
Page 38: ...3 2 ...
Page 46: ...4 2 ...
Page 92: ...5 34 ...
Page 94: ...6 2 ...
Page 98: ...6 6 ...
Page 100: ...7 2 ...
Page 112: ...7 14 ...
Page 114: ...8 2 ...
Page 120: ...8 8 ...
Page 122: ...9 2 ...
Page 134: ...9 14 ...
Page 136: ...10 2 ...
Page 144: ...10 10 ...
Page 146: ...11 2 ...
Page 154: ...A1 2 ...
Page 158: ...A2 2 ...
Page 169: ...GENERAL TITRATION APPLICATIONS BROCHURE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 170: ......
Page 195: ...TITRATION THEORY AUTOMATIC POTENTIOMETRIC TITRATOR HI932 HI931 ...
Page 196: ...2 ...
Page 198: ...4 ...
Page 216: ...22 THEORY931 932 10 18 ...
Page 217: ......