7-5
pH MODE
7.2.2. FIRST CALIBRATION POINT
Option: Point or Offset
If
Point
is selected, the slope values adjacent to the calibration points will be reevaluated (normal calibration).
If
Offset
is selected the existing slope values will not be changed.
7.2.3. EDIT CUSTOM BUFFERS
If you wish to use buffers other than the standard ones, use the
Edit Custom Buffers
option to set the desired pH value, up to five pH
custom buffers can be set.
Note
: Custom buffers are not temperature compensated, enter the value of the buffer at the calibration temperature.
• Use
and
keys to select the desired buffer.
• Press
to delete the selected buffer.
• Press
to edit the selected buffer.
Summary of Contents for HI931
Page 1: ...MANUALS AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 2: ......
Page 4: ......
Page 5: ...QUICK START GUIDE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 9: ...5 QUICK START GUIDE TITRATOR CONNECTIONS FRONT VIEW REAR VIEW ...
Page 19: ...INSTRUCTION MANUAL AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 20: ......
Page 24: ...1 2 ...
Page 26: ...2 2 ...
Page 29: ...2 5 SETUP 2 3 2 TITRATOR REAR VIEW 2 3 3 TITRATOR REAR VIEW WITH PERISTALTIC PUMP ...
Page 36: ...2 12 ...
Page 38: ...3 2 ...
Page 46: ...4 2 ...
Page 92: ...5 34 ...
Page 94: ...6 2 ...
Page 98: ...6 6 ...
Page 100: ...7 2 ...
Page 112: ...7 14 ...
Page 114: ...8 2 ...
Page 120: ...8 8 ...
Page 122: ...9 2 ...
Page 134: ...9 14 ...
Page 136: ...10 2 ...
Page 144: ...10 10 ...
Page 146: ...11 2 ...
Page 154: ...A1 2 ...
Page 158: ...A2 2 ...
Page 169: ...GENERAL TITRATION APPLICATIONS BROCHURE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 170: ......
Page 195: ...TITRATION THEORY AUTOMATIC POTENTIOMETRIC TITRATOR HI932 HI931 ...
Page 196: ...2 ...
Page 198: ...4 ...
Page 216: ...22 THEORY931 932 10 18 ...
Page 217: ......