7-12
pH MODE
•
In ATC mode, the pH value for custom buffers can be modified by pressing
.
•
If the Automatic Buffer entry type was selected for the calibration procedure, the titrator will automatically select the closest
buffer to the measured pH value from the buffer group.
•
If the Semiautomatic Buffer entry type was selected use the
or
to select the buffer. Only buffers in the
buffer group will be displayed.
CALIBRATION MESSAGES:
•
Wrong Buffer. Please check the buffer:
This message appears when the difference between the pH reading and the value of
the selected calibration buffer is significant. If the message is displayed, check if you have selected the appropriate calibration
buffer.
• Wrong buffer temperature:
This message appears if the buffer temperature is out of the defined temperature range.
•
Clean the electrode or check the buffer. Press
to update calibration:
This message alerts the user that some dirt
or deposits could be on the electrode, or the buffer is contaminated.
• Slope too low. Please check the buffer:
This message appears if the current slope is under 80% of the default slope.
Recalibrate the instrument using fresh buffers.
• Slope too high. Press
to clear the old calibration:
This message appears as a result of an erroneous slope
condition.
7.4. LOGGING
Data logging is available in pH mode. It can be logging on demand (Manual Logging) or automatically (Interval Logging) at
predefined time intervals.
To customize the logging report:
• Press
to display the Data Parameters screen.
• Highlight the
Setup pH/mV/ISE
Report option and press
to display the
Setup pH/mV/ISE
Report screen.
• Use the
and
keys to highlight the data field that you want to show/hide in the pH/mV/ISE report and then press
to activate/deactivate it.
• Each field marked by “*” is an active field selected for the report.
• Press
to save the customized report.
Summary of Contents for HI931
Page 1: ...MANUALS AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 2: ......
Page 4: ......
Page 5: ...QUICK START GUIDE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 9: ...5 QUICK START GUIDE TITRATOR CONNECTIONS FRONT VIEW REAR VIEW ...
Page 19: ...INSTRUCTION MANUAL AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 20: ......
Page 24: ...1 2 ...
Page 26: ...2 2 ...
Page 29: ...2 5 SETUP 2 3 2 TITRATOR REAR VIEW 2 3 3 TITRATOR REAR VIEW WITH PERISTALTIC PUMP ...
Page 36: ...2 12 ...
Page 38: ...3 2 ...
Page 46: ...4 2 ...
Page 92: ...5 34 ...
Page 94: ...6 2 ...
Page 98: ...6 6 ...
Page 100: ...7 2 ...
Page 112: ...7 14 ...
Page 114: ...8 2 ...
Page 120: ...8 8 ...
Page 122: ...9 2 ...
Page 134: ...9 14 ...
Page 136: ...10 2 ...
Page 144: ...10 10 ...
Page 146: ...11 2 ...
Page 154: ...A1 2 ...
Page 158: ...A2 2 ...
Page 169: ...GENERAL TITRATION APPLICATIONS BROCHURE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 170: ......
Page 195: ...TITRATION THEORY AUTOMATIC POTENTIOMETRIC TITRATOR HI932 HI931 ...
Page 196: ...2 ...
Page 198: ...4 ...
Page 216: ...22 THEORY931 932 10 18 ...
Page 217: ......