5-18
TITRA
TION METHODS
5.5.8. PRE-TITRATION VOLUME
During a titration, the equivalence point is reached after many titrant doses. These doses take up extra time while having no
relevance for equivalence point detection.
Pre-titration volume adds a large initial dose to jump directly to the proximity of the equivalence point.
This first dose occurs after the pre-titration stir time is completed.
The ranges for pre-titration volumes are shown below:
0.001 to 4.750 mL for a 5 mL burette
0.001 to 9.500 mL for a 10 mL burette
0.005 to 23.750 mL for a 25 mL burette
0.005 to 47.500 mL for a 50 mL burette
To disable a pre-titration volume, enter 0.000 mL.
Note:
A pre-titration volume is highly recommended whenever possible. Fewer doses will considerably shorten the overall titration
duration.
5.5.9. PRE-TITRATION STIR TIME
Option: 0 to 180 seconds
When enabled, the sample is mixed for a set period of time before any titrant is added. This allows the sample to become
homogeneous.
The
Pre-Titration Stir Time
option is disabled if 0 seconds is entered.
Summary of Contents for HI931
Page 1: ...MANUALS AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 2: ......
Page 4: ......
Page 5: ...QUICK START GUIDE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 9: ...5 QUICK START GUIDE TITRATOR CONNECTIONS FRONT VIEW REAR VIEW ...
Page 19: ...INSTRUCTION MANUAL AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 20: ......
Page 24: ...1 2 ...
Page 26: ...2 2 ...
Page 29: ...2 5 SETUP 2 3 2 TITRATOR REAR VIEW 2 3 3 TITRATOR REAR VIEW WITH PERISTALTIC PUMP ...
Page 36: ...2 12 ...
Page 38: ...3 2 ...
Page 46: ...4 2 ...
Page 92: ...5 34 ...
Page 94: ...6 2 ...
Page 98: ...6 6 ...
Page 100: ...7 2 ...
Page 112: ...7 14 ...
Page 114: ...8 2 ...
Page 120: ...8 8 ...
Page 122: ...9 2 ...
Page 134: ...9 14 ...
Page 136: ...10 2 ...
Page 144: ...10 10 ...
Page 146: ...11 2 ...
Page 154: ...A1 2 ...
Page 158: ...A2 2 ...
Page 169: ...GENERAL TITRATION APPLICATIONS BROCHURE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 170: ......
Page 195: ...TITRATION THEORY AUTOMATIC POTENTIOMETRIC TITRATOR HI932 HI931 ...
Page 196: ...2 ...
Page 198: ...4 ...
Page 216: ...22 THEORY931 932 10 18 ...
Page 217: ......