9
QUICK ST
ART GUIDE
Setup Titration Report
Users can select the information that is stored for each titration.
To setup the titration report, follow the procedure below:
• From the main screen, press
. The
Data Parameters
screen will be displayed.
• Highlight
Setup Titration Report
and press
.
• Mark the fields to be included in the titration report with the “*” symbol. Use the
and
keys to highlight a field and
/
to toggle the field.
• Press
to save the customized report .
Preparing the Sample
• Add 50 to 65 mL of distilled/deionized water to the titration beaker.
• Use a pipette or burette to add 5.0 mL of the sample (0.1M Hydrochloric Acid (HCl)) into the same beaker.
• Slide the stirrer assembly up.
• Place the beaker under the stirrer assembly.
• Lower the stirrer assembly until the electrodes are submersed and the stirrer is close to the bottom of the beaker.
• Adjust the level of the sample solution with distilled/deionized water so that the pH electrode bulb is completely immersed in
the sample solution and the reference junction of the electrode is 5-6 mm below the surface.
Performing a Titration
• From the main screen, press
. You will be prompted to enter the analyte size. Enter 5 mL and press
. The titrator will
start the analysis.
• At the end of the titration, the message “Titration Completed” will appear on the display with the final concentration of the
analyte in the sample and the equivalence endpoint volume.
Understanding the Displayed Information
During a titration the following screen is displayed:
Summary of Contents for HI931
Page 1: ...MANUALS AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 2: ......
Page 4: ......
Page 5: ...QUICK START GUIDE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 9: ...5 QUICK START GUIDE TITRATOR CONNECTIONS FRONT VIEW REAR VIEW ...
Page 19: ...INSTRUCTION MANUAL AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 20: ......
Page 24: ...1 2 ...
Page 26: ...2 2 ...
Page 29: ...2 5 SETUP 2 3 2 TITRATOR REAR VIEW 2 3 3 TITRATOR REAR VIEW WITH PERISTALTIC PUMP ...
Page 36: ...2 12 ...
Page 38: ...3 2 ...
Page 46: ...4 2 ...
Page 92: ...5 34 ...
Page 94: ...6 2 ...
Page 98: ...6 6 ...
Page 100: ...7 2 ...
Page 112: ...7 14 ...
Page 114: ...8 2 ...
Page 120: ...8 8 ...
Page 122: ...9 2 ...
Page 134: ...9 14 ...
Page 136: ...10 2 ...
Page 144: ...10 10 ...
Page 146: ...11 2 ...
Page 154: ...A1 2 ...
Page 158: ...A2 2 ...
Page 169: ...GENERAL TITRATION APPLICATIONS BROCHURE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 170: ......
Page 195: ...TITRATION THEORY AUTOMATIC POTENTIOMETRIC TITRATOR HI932 HI931 ...
Page 196: ...2 ...
Page 198: ...4 ...
Page 216: ...22 THEORY931 932 10 18 ...
Page 217: ......