7-6
pH MODE
• Use the numeric keypad to enter the pH buffer value.
• Press
to save the value.
• Press
to return to pH Setup menu.
7.2.4. EDIT BUFFER GROUP
Option: Up to five
Select up to five buffers from the available buffers (Hanna or Custom) to be used for automatic buffer recognition.
Within the buffer group, pH values must be at least 1.5 pH far apart.
If the buffer group already contains five pH buffers, at least one pH buffer has to be removed in order to add another buffer.
• Use the arrow keys to select the pH buffer to be included/removed in/from the buffer group.
• Press
or
to add/remove the selected pH buffer to/from buffer group.
• Press
to return to pH Setup menu.
Summary of Contents for HI931
Page 1: ...MANUALS AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 2: ......
Page 4: ......
Page 5: ...QUICK START GUIDE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 9: ...5 QUICK START GUIDE TITRATOR CONNECTIONS FRONT VIEW REAR VIEW ...
Page 19: ...INSTRUCTION MANUAL AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 20: ......
Page 24: ...1 2 ...
Page 26: ...2 2 ...
Page 29: ...2 5 SETUP 2 3 2 TITRATOR REAR VIEW 2 3 3 TITRATOR REAR VIEW WITH PERISTALTIC PUMP ...
Page 36: ...2 12 ...
Page 38: ...3 2 ...
Page 46: ...4 2 ...
Page 92: ...5 34 ...
Page 94: ...6 2 ...
Page 98: ...6 6 ...
Page 100: ...7 2 ...
Page 112: ...7 14 ...
Page 114: ...8 2 ...
Page 120: ...8 8 ...
Page 122: ...9 2 ...
Page 134: ...9 14 ...
Page 136: ...10 2 ...
Page 144: ...10 10 ...
Page 146: ...11 2 ...
Page 154: ...A1 2 ...
Page 158: ...A2 2 ...
Page 169: ...GENERAL TITRATION APPLICATIONS BROCHURE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 170: ......
Page 195: ...TITRATION THEORY AUTOMATIC POTENTIOMETRIC TITRATOR HI932 HI931 ...
Page 196: ...2 ...
Page 198: ...4 ...
Page 216: ...22 THEORY931 932 10 18 ...
Page 217: ......