5-32
TITRA
TION METHODS
5.5.21. VOLUME/FLOW RATE
The flow rate for the dosing system can be set by the user in an interval of 0.3 to two times the burette volume:
0.3 to 10 mL/min for a 5 mL burette
0.3 to 20 mL/min for a 10 mL burette
0.3 to 50 mL/min for a 25 mL burette
0.3 to 100 mL/min for a 50 mL burette
Note:
The titrator will automatically detect the burette size and display the correct high limit volume.
The flow rate is set for all burette operations.
5.5.22. SIGNAL AVERAGING
Option: 1,2,3 or 4 readings
This option enables filtering on the mV/pH reading.
If 1 Reading is selected, the filtering is disabled. The titrator will take the last reading and place it into a “moving window” along with
the last 2, 3 or 4 readings (depending on the selected option). The average of those readings is displayed and used for calculations.
Averaging more readings is helpful when a noisy signal is received from the electrode.
Summary of Contents for HI931
Page 1: ...MANUALS AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 2: ......
Page 4: ......
Page 5: ...QUICK START GUIDE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 9: ...5 QUICK START GUIDE TITRATOR CONNECTIONS FRONT VIEW REAR VIEW ...
Page 19: ...INSTRUCTION MANUAL AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 20: ......
Page 24: ...1 2 ...
Page 26: ...2 2 ...
Page 29: ...2 5 SETUP 2 3 2 TITRATOR REAR VIEW 2 3 3 TITRATOR REAR VIEW WITH PERISTALTIC PUMP ...
Page 36: ...2 12 ...
Page 38: ...3 2 ...
Page 46: ...4 2 ...
Page 92: ...5 34 ...
Page 94: ...6 2 ...
Page 98: ...6 6 ...
Page 100: ...7 2 ...
Page 112: ...7 14 ...
Page 114: ...8 2 ...
Page 120: ...8 8 ...
Page 122: ...9 2 ...
Page 134: ...9 14 ...
Page 136: ...10 2 ...
Page 144: ...10 10 ...
Page 146: ...11 2 ...
Page 154: ...A1 2 ...
Page 158: ...A2 2 ...
Page 169: ...GENERAL TITRATION APPLICATIONS BROCHURE AUTOMATIC POTENTIOMETRIC TITRATOR HI931 ...
Page 170: ......
Page 195: ...TITRATION THEORY AUTOMATIC POTENTIOMETRIC TITRATOR HI932 HI931 ...
Page 196: ...2 ...
Page 198: ...4 ...
Page 216: ...22 THEORY931 932 10 18 ...
Page 217: ......