APS-7000 Series User Manual
128
Term
Determines the step termination
settings at end of the step.
The CONT setting tells the
sequence to go to the next step.
The HOLD setting will pause the
output at the end of the step and
will only continue to the next step
when CONT[F3] is pressed.
The END setting will end the
sequence and go to Step 0(standby
step).
Trig Out
Sets the polarity for the trig out pin
on the J1 connector when the step
is running.
ON/OFF Phs Sets the start and stop phase of the
AC waveform. The ON Phs setting
sets the starting phase
of the step
.
OFF Phs sets the off phase
for the
output
when the output if turned
off.
Vset
Sets the AC voltage level. There are
4 secondary voltage settings that
determine how the voltage is
output.
Constant: Sets the voltage level of
the step immediately to Vset values.
Keep: Sets the voltage level to
“keep” the voltage of the previous
step.
Sweep: Linearly increases or
decreases the values from the end
of the previous step to the end of
the current step.
Summary of Contents for APS-7000 Series
Page 1: ...Programmable AC Power Source APS 7000 Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 111: ...OPERATION 111 Simulate Example ...
Page 157: ...COMMUNICATION INTERFACE 157 Exit 6 Press Exit F4 to exit from the serial port settings ...
Page 187: ...APPENDIX 187 APS 7000 Dimensions APS 7050 402 545 5 430 110 4 87 9 Scale mm ...
Page 188: ...APS 7000 Series User Manual 188 APS 7100 561 6 705 5 430 110 4 87 9 Scale mm ...
Page 189: ...APPENDIX 189 APS 7200 402 5 310 5 429 8 682 4 649 2 Scale mm ...
Page 190: ...APS 7000 Series User Manual 190 APS 7300 491 5 399 5 429 8 682 4 649 2 Scale mm ...