OPERATION
83
Triggering
Example:
Site selection: The surge/dip site is
selected as shown below.
Trigger (0
º for Auto
)
Dip/Surge site
T2
T1
ACV
100mS
ACV
Sets the ACV surge/dip level from
the 0V level.
T1
Sets the T1 time.
Range: 0~22ms(Auto Mode)
Range: 0~99ms(Manual Mode)
T2
Sets the width of the surge/dip.
Range: 0~22ms(Auto Mode)
Range: 0~99ms(Manual Mode)
100mS
Fixed 100mS delay after triggering.
Steps
1.
Press the
Menu
key. The Menu
setting will appear on the display.
Menu
2.
Use the scroll wheel to go to item 2,
Surge/Dip
Control
and press
Enter
. Alternatively, use the
short-cut key shift+8.
3.
Go to the
Mode
setting using the scroll wheel
and press
Enter.
Select the desired mode and
press
Enter
again to confirm.
The Manual mode will allow you to manually
trigger the surge/dip site. The Automatic
setting will automatically trigger the surge/dip
site.
Mode
Manual, Auto, OFF
Summary of Contents for APS-7000 Series
Page 1: ...Programmable AC Power Source APS 7000 Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 111: ...OPERATION 111 Simulate Example ...
Page 157: ...COMMUNICATION INTERFACE 157 Exit 6 Press Exit F4 to exit from the serial port settings ...
Page 187: ...APPENDIX 187 APS 7000 Dimensions APS 7050 402 545 5 430 110 4 87 9 Scale mm ...
Page 188: ...APS 7000 Series User Manual 188 APS 7100 561 6 705 5 430 110 4 87 9 Scale mm ...
Page 189: ...APPENDIX 189 APS 7200 402 5 310 5 429 8 682 4 649 2 Scale mm ...
Page 190: ...APS 7000 Series User Manual 190 APS 7300 491 5 399 5 429 8 682 4 649 2 Scale mm ...