OPERATION
73
Note
OC-FOLD can only be active when the I rms level is
greater than 0.
7.
Press
I
rms
and then press
OC-FOLD[F1]
toggle the OC-Fold function on or off.
OC-Fold setting
Setting the On/Off Phase
Background
The on phase setting sets the starting phase of
the voltage output. The off phase setting sets
the ending phase of the voltage output.
Steps
1.
Press
Shift + 7
or
Shift + 4
the On Phase or Off
Phase, respectively.
Shift
+
7
On Phase
or
4
Off Phase
2.
Set the
On Phase
or
Off Phase
setting with the
scroll wheel/keypad or with the F3 ~ F4 soft-
keys. The MAX and MIN soft-keys set the
phase to the maximum and minimum,
respectively.
Range
0 ~ 359
º
Soft-keys
MAX, MIN