APS-7000 Series User Manual
68
1.
Press
Shift + I
rms
and then press
DELAY[F2]
.
2.
Set the desired delay time with the scroll
wheel/keypad or with the F3 ~ F4 soft-keys.
The MAX and MIN soft-keys set the delay
time to the maximum and minimum,
respectively.
Range
0(off) ~ 10 seconds
Soft-keys
MAX, MIN
3.
Press
Enter
to confirm the delay time setting.
Example
Delay Time
Min/Max
settings
F3
F4
IPK Measurement
Range Settings
The IPK Range settings allow you to manually set
the peak current measurement range. By default
this setting is set to AUTO.
1.
Press Shift + I rms and then press IPK Range[F1].
2.
Set the desired range with the scroll wheel.
Range
AUTO, 0.28A, 1.4A, 14A, 70A
(APS-7050, APS-7100)
AUTO, 14A, 140A
(APS-7200, APS-7300)
3.
Press
Enter
to confirm the IPK Range setting.