OPERATION
79
Using the remote sense (APS-7200 and APS-7300 only)
The APS-7200 and APS-7300 can be operated using local or remote
voltage sense. By default, the power supply is configured for local
sense.
Warning
Ensure the output is off before handling the
remote sense connectors.
Use sense cables with a voltage rating exceeding
the isolation voltage of the power supply.
Never connect sensing cables when the output is
on. Electric shock or damage to the power supply
could result.
Remote sense
connectors
overview
The remote sense connectors
are located at the rear panel of
the APS-7200 and APS-7300.
Local Sense
Local sense
operation
When using local sense, the sensing terminals
are not used. No compensation of any possible
voltage drop seen on the load cables is
performed. Local sense is only recommended
when the voltage drop is of no consequence. By
default, the power supply is configured for
local sense.
1.
Check that the remote sense setting is disabled
(page 93).
Summary of Contents for APS-7000 Series
Page 1: ...Programmable AC Power Source APS 7000 Series USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Page 111: ...OPERATION 111 Simulate Example ...
Page 157: ...COMMUNICATION INTERFACE 157 Exit 6 Press Exit F4 to exit from the serial port settings ...
Page 187: ...APPENDIX 187 APS 7000 Dimensions APS 7050 402 545 5 430 110 4 87 9 Scale mm ...
Page 188: ...APS 7000 Series User Manual 188 APS 7100 561 6 705 5 430 110 4 87 9 Scale mm ...
Page 189: ...APPENDIX 189 APS 7200 402 5 310 5 429 8 682 4 649 2 Scale mm ...
Page 190: ...APS 7000 Series User Manual 190 APS 7300 491 5 399 5 429 8 682 4 649 2 Scale mm ...