COMMUNICATION INTERFACE
163
Function Active
ON, OFF
6.
Set the GPIB address.
GPIB Address
0 ~ 30
GPIB port configuration
Exit
7.
Press
Exit[F4]
to exit from the
serial port settings.
GPIB constraints
Maximum 15 devices altogether, 20m cable
length, 2m between each device
Unique address assigned to each device
At least 2/3 of the devices turned On
No loop or parallel connection
GPIB Function Check
Functionality
check
Please use the National Instruments
Measurement & Automation Controller
software to confirm GPIB/LAN functionality.
See the National Instrument website,
http://www.ni.com for details.
Note
For further details, please see the programming
manual, available on the GW Instek web site @
www.gwinstek.com.