APS-7000 Series User Manual
62
2.
Set the voltage limit with the scroll wheel or
with the F3 ~ F4 soft-keys. The MAX and MIN
soft-keys set the limit to the maximum and
minimum, respectively.
Range
10% of full range ~ full range
Soft-keys
MAX, MIN
3.
Press
Enter
to confirm the voltage limit setting.
Volt Limit
Min/Max
settings
F3
F4
Note
Each voltage range (155V, 310V, 600V) has an
independent voltage limit.
Before change volt limit setting, if ACV setting
value is bigger than desire volt limit value, so that
the volt limit value can't be change.
Setting the Output Voltage
The voltage setting sets the voltage level of the power supply.
Background
Before setting the power supply voltage level,
set the voltage range and voltage limit.
Steps
1.
Press the V key. The ACV
parameter will be editable.
V
V-Limit
2.
Set the voltage with the scroll wheel/keypad
or with the F1 ~ F4 soft-keys.