Figure 4.8: Analog Output Configuration Submenu
Four
editable parameters exist for each isolated Analog
Output. They are shown in Table 4.6.
Parameter
Meaning
Selectable Values
MODE
Analog Output
Signal Type
(Current/Voltage)
Current 4-20 (mA)
Voltage 2-10 (V DC)
Current 0-20 (mA)
Voltage 0-10 (V DC)
MEASURE
Measured
Parameter
assigned to the
Analog Output
None
Density
Viscosity
Temp Probe (Transducer
Sensor Tip Temp)
Temp Dome (Transducer
Dome (Block) Temp
Temp Elec (Transmitter
Enclosure Temp)
MIN
Low alarm value
–
any value
below this
threshold will
trigger an Alarm
state
Value entered using the
keypad.
MAX
High alarm value
–
any value
above this
threshold will
trigger an Alarm
state
Value entered using the
keypad.
Table 4.6: Analog Output Configuration Submenu
4.5.1.2 Editing an Isolated Analog Output
Configuration using the Keypad
•
From the Analog Output submenu of interest, use the
Up/Down arrow keys to scroll through the editable
parameters.
•
Press
ENTER
to select the desired parameter.
•
Edit the choice/value for that parameter. Enter numeric
values using the keypad as necessary.