ViscoSite VL800 Viscometer
Operating Manual
ViscoSite VL800 Viscometer
Galvanic Applied Sciences, Inc.
Email: [email protected]
Phone: (978) 848--2701
Operating Manual Rev: 2.0E
Toll-Free: 1 (866) 252-8470
Page 71 of 132
The COM Setup submenu, used to configure the ViscoSite
transmitter’s serial port, is shown in Figure 4.12.
Figure 4.12: COM Setup Submenu
The serial port communications protocol options are RS-
232 and RS485.
4.6.4.2 Editing a COM Setup Submenu Parameter
using the Keypad
•
Use the Up/Down arrow keys to scroll through the
Options entries to the Com Setup submenu.
•
Press
ENTER
to select COM Setup.
•
Use the Up/Down arrow keys to scroll through the
entries to the desired serial communications parameter.
•
Press
ENTER
to select and save the edited value to the
transmitter.
Pressing
ESC
during COM Setup when the cursor is at the
end of a line aborts parameter editing without saving
changes.
Pressing
ESC
from the COM Setup submenu when NOT
actively editing a parameter returns to the main Option
menu. Pressing
ESC
again returns to the Data Display.
Power cycle the transmitter to complete the configuration
of the COM communications protocol.
Changes to the COM Port configuration do not take effect
until the transmitter is power cycled.
The COM setup must be configured via the transmitter
keypad.