ViscoSite VL800 Viscometer
Operating Manual
ViscoSite VL800 Viscometer
Galvanic Applied Sciences, Inc.
Email: [email protected]
Phone: (978) 848--2701
Operating Manual Rev: 2.0E
Toll-Free: 1 (866) 252-8470
Page 51 of 132
saved to the transmitter via the transmitter keypad. See
Section 4.6.3.
•
Ensure
the transmitter’s LAN port is connected to the
LAN.
•
Connect the PC’s Ethernet LAN cable to the LAN on
which the ViscoSite transmitter resides.
•
See 5 for instructions regarding initiation of a
communications session between the
remote
PC
running the ViscoSite software and the ViscoSite
transmitter.
Simultaneous Communication Sessions – USB
and Ethernet
When a ViscoSite transmitter is simultaneously connected both
locally
via
USB
and
remotely
via
Ethernet
,
commands from the Ethernet-
connected remote computer are locked out
until the USB connection
is broken.
Once the USB connection is broken, the commands from the
remote
computer via the
Ethernet
port remain locked out until no transmitter
keypad activity has occurred for 3 minutes.
There is
no lockout
between the USB
and the transmitter’s keypad
(see 4.) Both are
local
connections. Parameter changes
can
be made via
the transmitter keypad while a
local
USB connection is present.