Figure 3.10: Standard IS Barrier Connection General
Applications (Up to 200 °C)
Figure 3.11: IS Barrier Connections - High Temperature
Applications (200 °C to 450 °C). (Includes Transducer Block
(Dome) RTD)
3.7.3.2 Grounding of the Intrinsically Safe Barriers
The Intrinsically Safe barriers, as installed in the transmitter
enclosure from the factory, are connected to earth ground
via
the barrier mounting rail’s attachment to the chassis
.
The Intrinsically Safe barriers
MUST
be connected to earth
ground to provide a safe transducer installation in the
hazardous area.
Failure to ground the IS barriers MAY result in shock,
injury, fire, explosion, and / or death.
If the IS barriers must be relocated outside the transmitter
enclosure, the Customer assumes all associated costs and
Intrinsically Safe
Side (Transducer in
Hazardous Area)
Non-Intrinsically Safe Side
(Transmitter in Non-
Hazardous / Declassified Area)
Intrinsically Safe
Side (Transducer in
Hazardous Area)
Non-Intrinsically Safe Side
(Transmitter in Non-
Hazardous / Declassified Area)