The Alarm Source column cannot be edited. It contains a simple list of
all possible parameters than can be configured to display an alarm on
the LCD.
The values in the “Units” column are derived from the measurement
units selected in the Display Units portion of the Display Setup screen.
See Section 5.6.2 to change the Display Units if necessary.
The Alarm DISPLAY limits configured here apply ONLY to the value of
the selected alarmed parameter that is DISPLAYED on the LCD. They
do NOT correspond to the limits associated with any analog or relay
outputs that may also be configured to output based on the value of
the same parameter.
This allows the user to set a tighter parameter value range to trigger a
VISUAL DISPLAY alarm
on a parameter that is outside ideal limits but
not far enough from ideal to warrant an analog output or relay (that has
been configured to monitor the same parameter) going into full alarm
status.
5.6.3.1 Selecting and Editing a Display Alarm
Configuration
•
Double click on the LCD display line on which the alarm
is to appear.
•
Select/left click the desired display line number.
•
Double click on the parameter column on the same line
in the table to display a drop-down list of the
parameters available for alarm display on the LCD.
•
Select the desired parameter from the drop-down list.
•
Press
ENTER/RETURN
on the keyboard to select the
parameter for alarm display.
•
Make additional parameter selections as necessary,
pressing ENTER/RETURN after each selection
.
•
Once all parameters have been configured in this
submenu as described above,
click on another toolbar
button to trigger the “Save Changes”
Dialog
.
•
To save the most recent changes, click “
Save
.”
This
sends the updated information to the transmitter.
•
To exit the menu without saving the most recent
changes
, click “
Continue
.”
•
To remai
n in the menu, click “
Cancel
.” The most recent
changes remain on the screen for further editing. They