ViscoSite VL800 Viscometer
Operating Manual
ViscoSite VL800 Viscometer
Galvanic Applied Sciences, Inc.
Email: [email protected]
Phone: (978) 848--2701
Operating Manual Rev: 2.0E
Toll-Free: 1 (866) 252-8470
Page 70 of 132
Subnet
–
the Subnet mask for the network
The transmitter must be assigned a
static
IP address.
The ViscoSite system will not negotiate a connection with a
DHCP server.
The network configuration values should be selected in
consultation with the IT department at the user’s facility.
Incorrect setting of these values will result in the transmitter
being unable to communicate with the Remote PC via the
LAN and COULD cause network-wide disturbances which
could affect other plant instrumentation.
The Network Configuration parameters can only be entered
via the ViscoSite transmitter keypad.
4.6.3.2 Editing a Network Configuration Submenu
Parameter using the Keypad
•
From the Option submenu, Use the Up/Down arrow
keys to scroll through the entries to the Network Config
submenu.
•
Press
ENTER
to select Network Config.
•
Edit the numeric parameters. Include decimal points as
required.
•
Press
ENTER
after editing each parameter to save the
edited value to the transmitter after entry.
Pressing
ESC
during network configuration editing when
the cursor is at the end of a line aborts parameter editing
without saving changes.
Pressing
ESC
from the Network Config submenu when NOT
actively editing a parameter returns to the main Option
menu. Pressing
ESC
again returns to the Data Display.
COM Setup
4.6.4.1 Accessing the COM Setup Submenu
•
From the Data Display, press the Option Setup button
on the keypad.
•
The Option submenu shown in Figure 4.10 is displayed.
•
Use the Up/Down arrow keys to scroll through the
Option entries to the COM Setup submenu.