Parameter Meaning
Selectable Values
MEASURE
Parameter set to
alarm on the given
relay
None
Density
Viscosity
Temp Probe
(Transducer Sensor
Tip Temperature)
Temp Dome
(Transducer-block RTD
temperature)
Temp Elec
(Transmitter Enclosure
Temperature)
MIN
Low alarm value
–
any value below
this threshold will
trigger the relay
alarm behavior
Value entered using
the keypad.
MAX
High alarm value
–
any value above
this threshold will
trigger the relay
alarm behavior
Value entered using
the keypad.
Table 4.7: Relay Configuration Submenu
4.5.1.4 Editing a Relay Output Configuration Using
the Keypad
•
From the Relay Output submenu of interest, use the
Up/Down arrow keys to scroll through the three editable
parameters.
•
Press
ENTER
to select the parameter.
•
Edit the value for that parameter. Enter numeric values
using the keypad.
•
Press
ENTER
to save the edited value of the selection to
the transmitter.
Pressing
ESC
during relay output configuration editing
when the cursor is at the end of a line aborts parameter
editing without saving changes.
Pressing
ESC
from a Relay Output submenu when NOT
actively editing a parameter returns to the main Output
Configuration menu. Pressing
ESC
again returns to the Data
Display.