ViscoSite VL800 Viscometer
Operating Manual
ViscoSite VL800 Viscometer
Galvanic Applied Sciences, Inc.
Email: [email protected]
Phone: (978) 848--2701
Operating Manual Rev: 2.0E
Toll-Free: 1 (866) 252-8470
Page 96 of 132
All temperature values must be input in degrees Celsius.
All viscosities must be entered in the same units as those
displayed on the LCD.
•
Click in the
Normal Temperature
box and enter the
expected
“normal”
material temperature the probe is
expected to see when the process is functioning
“ideally” or “normally.”
•
Click
ENTER/RETURN
on the keyboard.
•
Click in the
Low Temperature
box and enter the lowest
material temperature the transducer probe is expected
to see when the process is functioning normally.
•
Click
ENTER/RETURN
on the keyboard.
•
Click in the
Viscosity
box to the right of the Low
Temperature box and enter the viscosity of the process
material at the temperature entered into the Low
Temperature box.
•
Click
ENTER/RETURN
on the keyboard.
•
Click in the
High Temperature
box and enter the highest
material temperature the transducer probe is expected
to see when the process is functioning normally.
•
Click
ENTER/RETURN
on the keyboard.
•
Click in the
Viscosity
box to the right of the High
Temperature box and enter the viscosity of the process
material at the temperature entered into the High
Temperature box.
•
Click
ENTER/RETURN
on the keyboard.
•
Click on another toolbar button to trigger the “
Save
Changes
” Dialog
.
•
To save the most recent changes, click “
Save
.”
This
sends the updated information to the transmitter.
•
To exit the menu without saving the most recent
changes
, click “
Continue
.”
•
To remain in the menu, click “
Cancel
.” The most recent
changes remain on the screen for further editing. They
are not sent to the transmitter.
5.8.4.2 Disabling Temperature Compensation
•
Set all parameter values outlined in Section 5.8.4.1 to
zero (0), clicking
ENTER/RETURN
on the keyboard after
each entry.
•
Click on another toolbar button to trigger the “
Save
Changes
” Dialog
.
•
To save the most recent changes, click “
Save
.”
This
sends the updated information to the transmitter.