ViscoSite VL800 Viscometer
Operating Manual
ViscoSite VL800 Viscometer
Galvanic Applied Sciences, Inc.
Email: [email protected]
Phone: (978) 848--2701
Operating Manual Rev: 2.0E
Toll-Free: 1 (866) 252-8470
Page 89 of 132
Parameter
Explanation
<Blank>
Nothing output on this line if this option is
selected
Viscosity
Present calculated viscosity parameter
Temp Probe
Present temperature measured by the transducer
probe RTD
Temp Dome
Present temperature measured by the transducer
block RTD
Temp Elec
Present temperature measured inside the
transmitter enclosure
Frequency
Present oscillation frequency of the transducer
Temp Comp Visc Present measured viscosity compensated for
sample temperature
Table 5.4: Parameters Available for Output on an Analog Channel
5.7.1.1 Selecting and Editing an Output Type and/or
Measurement Parameter for Analog Outputs
•
Double click on the desired field.
•
Select the desired value from the drop-down menu that
appears.
•
Press
ENTER/RETURN
on the keyboard.
•
Make additional parameter selections or entries as
necessary,
pressing ENTER/RETURN after each
selection.
•
Once all parameters have been configured in this
submenu as described above,
click on another toolbar
button to trigger the “Save Changes” Dialog.
•
To save the most recent changes, click “
Save
.”
This
sends the updated information to the transmitter.
•
To exit the menu without saving the most recent
changes,
click “
Continue
.”
•
To remain in the menu, click “
Cancel
.” The most recent
changes remain on the screen for further editing. They
are not sent to the transmitter.
5.7.1.2 To Change/Edit Min and/or Max Numerical
Values for Analog Outputs
•
Click on the desired cell.
•
Type in the new value.
•
Press
ENTER/RETURN
on the keyboard.
•
Make additional parameter entries as necessary,
pressing ENTER/RETURN after each value entered.
•
Once all parameters have been configured in this
submenu as described above,
click on another toolbar
button to trigger the “Save Changes” Dialog.
•
To save the most recent changes, click “
Save
.”
This
sends the updated information to the transmitter.