![Fluke PG9000 Series Operation And Maintenance Manual Download Page 189](http://html1.mh-extra.com/html/fluke/pg9000-series/pg9000-series_operation-and-maintenance-manual_2309961189.webp)
Page 179
© 1998-2010 Fluke Calibration
6.
Troubleshooting
6.1
Overview
PG9000 is a sophisticated pressure measuring instrument with advanced on-board features
and functions. Before assuming that unexpected behavior is caused by a system defect or breakdown,
the operator should use this manual and other training facilities to become thoroughly familiar with
PG9000 operation. This troubleshooting guide is intended as an aid in identifying the cause of
unexpected behavior and determining whether the behavior is due to normal operation or an internal or
external problem.
Identify the symptom that you are observing from the
SYMPTOM
list below. A
PROBABLE CAUSE
is
provided and a
SOLUTION
is proposed including references to manual sections that provide information
that may be of assistance.
Table 25.
PG9000 Troubleshooting Checklist
SYMPTOM
PROBABLE CAUSE
SOLUTION
Will NOT power up.
Blown fuse.
Replace the power entry module fuse in the
PG9000 Terminal.
Front panel keys seem to be disabled. The host computer is remotely interfacing
with the PG9000.
Send “LOCAL” command from host
computer; or if using external software (such
as COMPASS for Pressure) stop the
program and press “ESC” on the PG
Terminal to return to local control; or cycle
PG9000 power. 0
Front panel display is dim.
Screen saver option has been activated.
Press any key to resume full screen power,
adjust activation time if desired. 3.12.4.1
Keypad presses make undesired
sounds or no sounds.
Keypad sound settings are incorrect.
Use sounds function to set keypad sounds
as desired. 3.12.4.2
Cannot access certain functions.
Display shows:
<ACCESS RESTRICTED>.
User Level restriction settings limit access
to critical functions. By default, editing
metrological information should be under
restricted access.
Change User Level or consult system
manager. 3.12.4.5
Cannot establish communication over
remote interface.
Computer and/or PG9000 interface is not
correctly configured; incorrect / mis-
matched port settings or bad interface
cable.
Check and correct interface configurations
and cables if necessary. Run COM port test.
3.12.5, 3.12.5.3, 4.2
Displays <FATAL ERROR> or
<FATAL FAULT>.
Encountered unresolved internal software
conflict.
Cycle power to clear. Please record
conditions leading up to the event, including
the numbers displayed when enter is
pressed and report the information to a Fluke
Calibration Authorized Service Provider.
Display <TOUT>, <TIME-OUT> or
<-TO->
PG9000 is having a communications
problem with an external barometer or
pressure generation/control component.
Check setup and communications with
external barometer or pressure
generation/control component. 3.12.5.4,
3.12.5.5
Displays <******> or <OVERFLOW>
where a numerical value should be.
Number to be displayed is too large for
allocated space. Usually due to an
erroneous setting or measurement
causing an out of limit high value to be
calculated.
Check settings that may be causing an out of
limit high measurement and adjust if
necessary. 3.10.3
The PG Terminal constantly displays
<Searching …. >.
The PG Terminal is unable to establish
communications with the PG9000 Platform
or REM.
Check that the 25-pin PG Terminal to REM
cable is installed correctly. If still unable to
get beyond <Searching …. >, contact a
Fluke Calibration Authorized Service
Provider. 2.3.1.1
Summary of Contents for PG9000 Series
Page 10: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page X Notes...
Page 128: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 118 Notes...
Page 164: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 154 Notes...
Page 188: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 178 Notes...
Page 192: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 182 Notes...