Select
<2ambT>
to specify the ambient temperature variable source for the temperature
values that are used to calculate ambient air density. Selecting
<2ambT>
offers the
temperature variable source choices:
<1internal>:
Use real time readings from PG9000’s ambient temperature platinum resistance
thermometer
in the Temperature - Humidity (TH) Probe
.
<2normal>:
Use a fixed value of 20
°
C.
<3user>:
Use a fixed user defined value. If
<3user>
is selected, the user value must be
entered, in degrees Centigrade (
°
C).
Select
<3%RH>
to specify the relative humidity variable source for the relative humidity
values that are used to calculate ambient air density. Selecting
<3%RH>
offers the relative
humidity variable source choices:
<1internal>:
Use real time readings from PG9000’s relative humidity sensor
in the
Temperature - Humidity (TH) Probe
.
<2normal>:
Use a fixed value of 45 %RH.
<3user>:
Use a fixed user defined value. If
<3user>
is selected, the user value must be
entered, in percent relative humidity (%RH).
Select
<4PCT>
to specify the piston-cylinder temperature variable source for the temperature
values that are used to compensate for thermal expansion of the piston-cylinder effective
area. Selecting
<4PCT>
offers the temperature variable source choices:
<1internal>:
Use real time readings from PG9000’s on-board piston-cylinder temperature
platinum resistance thermometers (embedded in the piston-cylinder module
mounting post).
<2normal>:
Use a fixed value of 20
°
C.
<3user>:
Use a fixed user defined value. If
<3user>
is selected, the user value must be
entered, in degrees Centigrade (
°
C).
Select
<5g>
to specify the value of acceleration due to gravity that is used in calculating the
force applied to the piston. Pressing
<5g>
offers the gravity variable source choices:
<1locall>:
Use the fixed value of local gravity stored under
[SPECIAL]
,
<6gl>
(see Section
<2normal>:
Use a fixed value of 9.806650 m/s
2
.
<3user>:
Use a fixed user defined value different from the current local gravity value
stored in the
[SPECIAL]
,
<6gl>
menu option. If
<3user>
is selected, the user
value must be entered in units of meters per second squared (m/s
2
).
Select
<6READY>
to edit the values of
Ready/Not Ready
status criteria (see Section 3.4).
These include:
<1Pposition>:
Piston position limits. Edit the fixed value, in millimeters (mm), of the band
around mid-float position within which the piston position is
Ready
(see
Section 3.4.1). This also determines the limit at which the piston is refloated by
the GEN function when GEN is ON (see Section 3.10.9).
<2vac>:
Maximum reference vacuum pressure when operating in absolute by vacuum
mode. Edit the value, in Pascal (Pa), under which the reference vacuum must
be for a vacuum
Ready
condition to occur (see Section 3.4.3).
Summary of Contents for PG9000 Series
Page 10: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page X Notes...
Page 128: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 118 Notes...
Page 164: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 154 Notes...
Page 188: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 178 Notes...
Page 192: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 182 Notes...