PG9602™ OPERATION AND MAINTENANCE MANUAL
© 2011 Fluke Calibration
Page
96
3.12.1.7 Edit a Mass Set
PURPOSE
To edit information contained in an existing mass set file. Use this function to
change mass set data after a recalibration or if a mass set configuration
changes.
PRINCIPLE & OPERATION
Caution
Mass set information is element specific metrological data.
Uninformed or accidental altering of this information may lead
to out of tolerance measurements. Mass set information should
only be edited by qualified personnel as part of the PG9000
calibration or recalibration process.
See Section 3.12.1.6 describing adding a new mass set. The principles and
procedures to add or edit a mass set are identical except that the mass set to
edit is selected from the existing mass sets screen.
The mass set type (manual or AMH) cannot be changed when a mass set is edited.
3.12.1.8 View a Mass Set
PURPOSE
To view the individual mass set details of an existing mass set.
OPERATION
To view information contained in an existing mass set file press
[SPECIAL]
,
<1PC/MS>
and select
<2mass set>
,
<1view>
. The viewing function operates in
the same manner as the add function (see Section 3.12.1.6).
3.12.1.9 Delete a Mass Set
PURPOSE
To delete an existing mass set file.
OPERATION
Caution
Once a mass set file has been deleted it cannot be recovered.
To recreate it, the add mass set function must be used (see
Section 3.12.1.6) which requires entering all the mass set
information. Before deleting a mass set, consider editing it (see
Section 3.12.1.7).
To delete a mass set press
[SPECIAL]
,
<1PC/MS>
and select
<2mass set>
,
<4delete>.
Select the mass set to be deleted and press
[ENT]
. Select
<2yes>
to confirm the mass set deletion. Press
<1no>
to NOT delete.
Note
The active mass set (mass set that is currently selected) cannot
be deleted. Before attempting to delete a mass bell, be sure it is
not currently selected.
3.12.1.10 Select Mass Set
PURPOSE
To select a mass set to be active and used by PG9000 in its calculation of the
reference pressure. To initialize the optional AMH automated if present.
Summary of Contents for PG9000 Series
Page 10: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page X Notes...
Page 128: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 118 Notes...
Page 164: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 154 Notes...
Page 188: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 178 Notes...
Page 192: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 182 Notes...