4. REMOTE OPERATION
Page 125
© 2011 Fluke Calibration
∗
ESE(?)
Purpose
Read or set the Standard Event Status Enable Register.
Syntax
“
∗
ESE
n
”
“
∗
ESE?”
Parameters
n:
’0 to 255’ This is the decimal representation of the bit(s) to
enable. To enable the PON and QYE bits, the argument would
be 128 + 4 = 132.
Query Reply
n
(0 to 255)
Remarks
The Standard Event Status Enable register determines which bits in the standard Event
Status Register are enabled and included in the Status Byte Register (ESB bit), and can
assert the SRQ line. The reply is in decimal numeric form.
Example
Command:
Reply:
“*ESE 132”
“OK” (using COM1. No reply if IEEE-488 port)
Command:
Reply:
“*ESE?”
“132”
∗
ESR?
Purpose
Read the Standard Event Register.
Syntax
“
∗
ESR?”
Query Reply
n
(0 to 255)
Remarks
The Standard Event Register contents are cleared after reading. The reply is in decimal
numeric form.
Example
Command:
Reply:
“*ESR?”
“4”
∗
IDN?
Purpose
Identify the PG9000 model, and serial number.
Syntax
“
∗
IDN?”
Remarks
The identification reply is made up of the manufacturer, the model, the serial number
and the software version. Each is separated by a comma.
Query Reply
The version string.
Example
Command:
Reply:
“*IDN?”
“FLUKE, PG9602, 1001,
Ver3.00a-Lib:nmcm
∗
OPC(?)
Purpose
Sets the operation complete bit when all operations have completed.
Syntax
“
∗
OPC”
“
∗
OPC?”
Remarks
This Command enables PG9000 to set the OPC bit in the Standard Event Status Register
when it has completed all pending functions. The Query replies with a “1” when all functions
are complete.
Since PG9000 does not support overlapping commands, this command has no practical use.
Query Reply
“0” or “1”
Example
Command:
Reply:
“*OPC”
“OK”
(using COM1. No reply if IEEE-488 port)
Command:
Reply:
“*OPC?”
“1”
∗
OPT?
Purpose
Reads the list of installed PG9000 options.
Syntax
“
∗
OPT?”
Remarks
This Query returns any registered option(s) installed in the PG. Each option is
separated by a comma.
Query Reply
A comma delimited text field of the installed options.
Example
Command:
Reply:
Reply:
“*OPT?”
“NONE” (no options installed)
“auto rotation” (auto rotation option)
∗
RST
Purpose
Resets the PG9000 settings to factory settings.
Syntax
“
∗
RST”
Remarks
This Command sets the PG9000 settings to factory settings which is equivalent to a
front panel executed
[SPECIAL], <5Reset>, 1sets
. This does not affect the
communications settings.
Example
Command:
Reply:
“*RST”
“OK” (using COM1. No reply if IEEE-488 port)
See Also:
Summary of Contents for PG9000 Series
Page 10: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page X Notes...
Page 128: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 118 Notes...
Page 164: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 154 Notes...
Page 188: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 178 Notes...
Page 192: ...PG9602 OPERATION AND MAINTENANCE MANUAL 2011 Fluke Calibration Page 182 Notes...